BECTON DICKINSON CATHENA; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 386801
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd cathena 18gx1.25in straight bc had foreign matter cathena cannula with substance inside tip of device.On retraction of stylet, customer noticed visible substance on inside of cannula near tip of device.
 
Additional Manufacturer Narrative
Our quality engineer inspected the two samples submitted for evaluation.The reported issue of foreign matter was confirmed upon inspection of the samples.Analysis of the samples showed no damage on the catheter tip.Foreign matter was observed on the cannula tip and covers.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records have been reviewed, and this batch meets our manufacturing specification requirements.Based on similar (b)(4), visible silicone was observed on the cannula, catheter and catheter body under 10x microscope.The assembly process of the cannula lubrication process was reviewed.The cannula is lubricated by dipping it into a silicone lube tank.During dipping, low air flow is blown through the cannula to prevent silicone flowing into the cannula.After dipping, the part is raised from the silicone lube tank followed by high pressure air blown through the cannula to remove the excess silicone.The catheter subassemblies are then set together with the needle hub subassembly.The assembled part then goes through a series of processes and loaded with the needle cover.There could have been silicone accumulated at the surface of the lube tank which might have transferred to the cannula surface during air blowing.After the catheter subassemblies are set together with the needle hub subassembly, given the silicone-like nature of the lube it is possible for the excess silicone to migrate to the cannula/catheter tip area during transportation or storage.To prevent this defect, revision of the procedure will be completed to include cleaning of the surface of the lube tank and surrounding areas every shift.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATHENA
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore NJ 07417
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22459392
Report Number2243072-2025-00838
Device Sequence Number12910872
Product Code FOZ
UDI-Device Identifier00382903868018
UDI-Public(01)00382903868018(11)241102(17)271031(10)4304881
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K220584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number386801
Device Lot Number4304881
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2025
Supplement Date Received by Manufacturer08/01/2025
Initial Report FDA Received Date07/09/2025
Supplement Report FDA Received Date08/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-