| Catalog Number |
6700625 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that a c+file readysteel 25mm 006 file broke during use.The broken part was not retrieved.Patient reported no discomfort; no special treatment was performed.Patient asked to report and discomfort and to follow up.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1821504).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (hand used file: technique no more verifiable to date), overuse (number of uses not communicated), excessive wear, patient condition and behaviour during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.
|
| |
|
Search Alerts/Recalls
|