| Catalog Number |
11-301343 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Joint Dislocation (2374)
|
| Date of Event |
03/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat: 110031011 lot: 66842409 28mm i.D.42mm o.D.Size e bearing, cat: 650-1157 lot: 3189493 delta cer fem hd 28/+3mm t1, g2: foreign ¿ australia , h6: suggested component code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided.Review of the available records identified the following: a superolateral dislocation.No fracture is identified.A definitive root cause cannot be determined.The complaint is confirmed based on the x-ray findings.Dislocation confirmed; periprosthetic fracture could not be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately two weeks post-implantation, the patient underwent a hip revision due to a periprosthetic fracture post-op and a dislocation as a result of the fracture.The bearing and head dislocated.There was internal fixation of the femur.Attempts have been made and no further information has been provided.
|
| |
|