• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1420
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventricular assist device (vad) patient presented to the emergency department (ed) after a controller fault alarm sounded and was admitted.The patient had no new symptoms or complaints.The controller was exchanged and the patient was discharged home.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Continuation d10: 6947m62 lead & e140 icd implant date: (b)(6) 2013 this information was received from the mechanical circulatory support product surveillance registry study.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for additional event details.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the controller was exchanged.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for additional information was received.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for investigation completion.Product event summary: the controller ((b)(6)) was not returned for evaluation.A review of the device history record was not performed as the reported event is not related to a manufacturing or servicing issue.Log file analysis could not be performed since log files were not available for analysis.As a result, the reported controller fault alarm event could not be confirmed.Applicable risk documentation and experience with events of similar circumstances were considered; a possible root cause of the reported controller fault alarm event may be attributed, but not limited, to a faulty internal component, memory corruption, and/or the patient allowing the batteries to deplete completely.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22461716
Report Number3007042319-2025-01433
Device Sequence Number8058671
Product Code DSQ
UDI-Device Identifier00888707006859
UDI-Public(01)00888707006859(17)211130
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2021
Device Model Number1420
Device Catalogue Number1420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer07/14/2025
07/24/2025
08/07/2025
Initial Report FDA Received Date07/09/2025
Supplement Report FDA Received Date07/23/2025
07/31/2025
08/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/19/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MCS1705PU VAD.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age44 YR
Patient SexMale
Patient Weight147 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-