An investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer experienced missing high/low glucose alarms with the reported application.Per the compatibility guide, use of the ios operating system 18.5 is incompatible with the reported application software version 2.12.0.8319.This guide is available to the user on the websites where the product is launched.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.Section d.Suspected medical device and g4 - pma/510(k)# has been populated for the freestyle librelink ios application as this report concerns a united kingdom customer.This is same/similar to us freestyle libre 2 ios application, model number 71926-01.The operating system is unknown so product information provided is for ios.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|