| Catalog Number |
ASC4630-01 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
06/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a meniscus repair surgery, the tristar 50 wand was used without a suction device.After the surgery, which was completed with the same device, it was found that the patient's skin had suffered first and second-degree burns.The burns were treated applying pressure to prevent the formation of blisters.The next day there were blisters that were removed.There was no surgical delay reported.The patient is currently recovering from the scalds.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).B1 and h6 updated.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A clinical review states that following surgery, the patient presented a first-degree burns, as shown in photo 1 labeled "day of surgery".The next day, the burns had progressed to second-degree, as seen in photo 2 labeled "day after surgery".This adverse event was likely procedure related, and the device had no influence on the event.As the ifu for the arthrowand does warn; ¿for wands with suction, failure to attach the suction adapter may cause thermal injury to the physician or patient.¿ please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for further assessment.
|
| |
|