• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number ETLW1620C124E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 06/17/2025
Type of Reportable Event Serious Injury
Event or Problem Description
An endurant iis stent graft system was implanted during the endovascular treatment of a 57 mm abdominal aortic aneurysm on (b)(6) 2025.It was reported that during the index procedure, the bifurcate stent graft landed a little lower then planned, therefore the physician opted to implanted an endurant ii aortic extension to cover up to the lowest renal artery.It was noted that anatomy played a role in the inaccurate delivery as the neck was conical and also the physician may have let the stent graft come down a little bit during the deployment process.Several hours after the procedure, the patient developed a cold leg and it was noted that the left iliac limb had thrombosed.It was confirmed the left limbs were 16-20/124, 16-13/124, and 16-10/124 that were occluded intervention was performed one day post the index procedure.Bare metal stents and a 16-16/124 endurant iliac limb were used to re-line the limb, and a non-mdt catheter was used to remove the clot.The healthcare professional indicated that the cause of the event was over-sizing, the physician felt the 16- 10 limb may have been slightly oversized.No additional clinical sequelae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: e tlw1610c124e, serial/lot #: (b)(6), ubd: 30-oct-2026, udi#: (b)(4); product id: etlw1613c124e, serial/lot #: (b)(6), ubd: 21-nov-2026, udi#: (b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key22466580
Report Number9612164-2025-03380
Device Sequence Number20818418
Product Code MIH
UDI-Device Identifier00643169467552
UDI-Public(01)00643169467552(17)270213
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberETLW1620C124E
Device Catalogue NumberETLW1620C124E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2025
Initial Report FDA Received Date07/10/2025
Date Device Manufactured02/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age70 YR
Patient SexMale
-
-