ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LUXOR OPHTHALMIC MICROSCOPES; MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC Back to Search Results
Model Number LX3
Medical Device Problem Codes Defective Component (2292); Unintended Movement (3026)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that the ophthalmic microscope's camera mount was damaged, and the microscope head handles were loose.Procedure details and patient impact have not been provided.Additional information has been requested; none has been received till date.
 
Additional Manufacturer Narrative
Additional information is provided in sections b.5 and h.6.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received reporting that the event occurred before cataract surgery.The surgery was completed on the same day.
 
Additional Manufacturer Narrative
Additional information is provided in sections d.9, h.3, h.6 and h.11.The company representative was able to confirm the reported events (damaged camera mount and loose handles).To address these issues, the nonconforming video adapter was replaced, and the handle setscrews were secured back into place.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is attributed to the nonconforming video adapter and unsecured handle setscrews.How are when the handle setscrews became unsecured is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUXOR OPHTHALMIC MICROSCOPES
Common Device Name
MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22466955
Report Number2028159-2025-01007
Device Sequence Number4106622
Product Code HRM
UDI-Device Identifier00380657530731
UDI-Public(01)00380657530731(17)331028(11)231031(10)16A01T
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLX3
Device Catalogue Number8065753073
Device Lot Number16A01T
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer07/15/2025
09/09/2025
Initial Report FDA Received Date07/10/2025
Supplement Report FDA Received Date08/04/2025
09/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-