| Catalog Number |
SRM-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed an opening assessed as fold flaw opening.As per the investigation procedure, creases, wear abrasion and non-penetrating nick were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|