• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC BAKRI TAMPONADE BALLOON CATHETER; OQY INTRAUTERINE TAMPONADE BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC BAKRI TAMPONADE BALLOON CATHETER; OQY INTRAUTERINE TAMPONADE BALLOON Back to Search Results
Model Number G30673
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A5: ethnicity: chinese, e1: phone number: (b)(6), h3: device evaluation is anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, following a cesarean delivery, the bakri tamponade balloon catheter's balloon ruptured during treatment of postpartum hemorrhage (pph) secondary to uterine atony.The patient experienced an estimated blood loss of 600ml.The balloon was then placed into the patient, but it did not inflate after several water injections.Upon inspection, the balloon was found to be ruptured.It was immediately replaced with a new one.The total estimated blood loss was 680ml.The patient did not require any blood transfusions.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.As reported, following a cesarean delivery, the bakri tamponade balloon catheter's balloon ruptured during treatment of postpartum hemorrhage (pph) secondary to uterine atony.The patient experienced an estimated blood loss of 600ml.The balloon was then placed into the patient, but it did not inflate after several water injections.Upon inspection, the balloon was found to be ruptured.It was immediately replaced with a new one.The total estimated blood loss was 680ml.The patient did not require any blood transfusions.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control (qc) procedures, and instructions for use (ifu), as well as a visual inspection of the returned device, were conducted during the investigation.Device returned without package or label.Upon visual inspection, it was confirmed that the balloon was ruptured.Cook has concluded that the device was manufactured to specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances that could have contributed to the reported failure mode.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there were no related non-conformances, adequate inspection activities had been established, there was objective evidence that the dhr was fully executed, and no other lot related complaints had been received from the field, it was concluded that there was no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling.The ifu packaged with the device contains the following in relation to the reported failure mode: how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." based on the available information, inspection of returned device, and the results of the investigation, cook has concluded that the nature and cause of the issue could not be determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAKRI TAMPONADE BALLOON CATHETER
Common Device Name
OQY INTRAUTERINE TAMPONADE BALLOON
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key22470188
Report Number1820334-2025-00841
Device Sequence Number18127775
Product Code OQY
UDI-Device Identifier10827002306735
UDI-Public(01)10827002306735(17)270612(30)1(10)16104894
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K170622
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG30673
Device Catalogue NumberJ-SOS-100500
Device Lot Number16104894
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer07/31/2025
Initial Report FDA Received Date07/10/2025
Supplement Report FDA Received Date08/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
Patient Weight75 KG
-
-