Model Number CL-936
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, four sutures broke while suturing at the connection of the needle and thread.Additional new sutures were used without issue.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
D10 concomitant products: cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36 (lot#:a5c1940y), cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36 (lot#:a5c1940y), cl-936, cl-936 polysorb* 1 vio 90cm gs12 x36 (lot#:a5c1940y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: b5, d9 (latest returned date), g3, h3, h6.H3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the account showed a breather pouch, four needles, and four sutures.The pouch was identified as a cl-936 polysorb suture, with all needles disengaged from their sutures.Inspection of the returned product revealed one suture and one needle, with the needle also disengaged.Microscopic examination showed instrument marks near the swaged end of the needle, and suture material was present in the swage, indicating breakage at that point.Functional testing was not performed as the needle was already detached upon return.It was reported that four sutures broke while suturing at the connection of the needle and thread.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a total knee replacement procedure, four sutures broke while suturing at the connection of the needle and thread.Additional new sutures were used without issue.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22470229
Report Number1219930-2025-02908
Device Sequence Number18139104
Product Code GAM
UDI-Device Identifier10884521043527
UDI-Public(01)10884521043527(17)300228(10)A5C1940Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCL-936
Device Catalogue NumberCL-936
Device Lot NumberA5C1940Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2025
Supplement Date Received by Manufacturer06/18/2025
08/20/2025
Initial Report FDA Received Date07/10/2025
Supplement Report FDA Received Date08/15/2025
09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11