• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP Back to Search Results
Model Number INFORNG
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump was returned to mmd for evaluation.A dhr review was completed and no non-conformances were found.During the investigation, mmd found that the pump pcb board had failed, which caused the no flow out alarm to fail.The pump was serviced to correct the issue.
 
Event or Problem Description
The initial reporter stated that the pump had broken hinges and a cracked door latch.When the pump was returned to mmd for evaluation, the no flow out alarm did not operate as expected.It failed to alarm.The initial reporter stated that the complaint had occurred during testing and had not had any adverse effect on a patient.The alarm failure was discovered at moog.[(b)(4)].
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ORANGE ENTERAL FEEDING PUMP
Common Device Name
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin egelhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key22470799
Report Number1722139-2025-00401
Device Sequence Number4106742
Product Code LZH
UDI-Device Identifier10814844000334
UDI-Public10814844000334
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K031199
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberINFORNG
Device Catalogue NumberINFORNG
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Initial Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-