| Catalog Number |
ECR45W |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 07/10/2025 d4: batch #unk.Additional information was requested, and the following was obtained: was there any other issue noted with the staple line other than bleeding (malformed staples, staple line missing staples, etc.)? malformed staples how was the bleeding controlled? suture sewing how much blood was lost (ml)? unknown did the patient require a transfusion? no is the current patient status known? discharged was there any patient consequence or change in the post-operative care of the patient as a result of the event? no a manufacturing record evaluation was performed for the finished device ecr45w with lot number 963a39; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the device was used to disconnect the left inferior pulmonary vein.After disconnecting the left inferior pulmonary vein, it was found that the nail edge at the distal end of the vein was not fully closed, resulting in bleeding.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 09/06/2025 an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|
|
|