| Catalog Number |
ECR45W |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic lobectomy procedure, it was observed that the staples were malformed after firing.They changed to another one to continue the procedure but the same problem happened again.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 07/10/2025.D4: batch #unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr45w with lot number 137c67; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 07/27/2025 d4: batch #872a74 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr45w reload (a) was received partially fired 2/3 and an open package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review = > a manufacturing record evaluation was performed for the finished device batch number 872a74, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|