| Model Number |
A1 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The mixer was replaced to resolve the issue.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a malfunction resulting in insufficient oxygen or fresh gas flow and the device did not automatically switch to alt o2.There was no patient involvement.
|
| |
|