MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TIBIA FIXED CEMENTED LEFT SIZE D STEM EXTENSION USE REQUIRED; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TIBIA FIXED CEMENTED LEFT SIZE D STEM EXTENSION USE REQUIRED; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42542006701
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Muscle Weakness/Atrophy (1967); Pain (1994); Osteopenia/ Osteoporosis (2651); Bursitis (4832); Joint effusion (4849); Aseptic loosening (4883); Tendinopathy (4887)
Date of Event 04/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via clinical study that underwent a total knee procedure.Subsequently, two (2) years and five (5) months post-implantation, the patient reported severe pain and a concern of weakness.Ongoing tendinitis/bursitis and mild effusion were also noted on exam.On the next follow-up six (6) months later, radiographic imaging displayed possible loosening and bone loss.The patient wishes to manage these symptoms conservatively with home care and observation at this time.All components remain implanted.No further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: medical product: persona revision fem cmt ccr pls sz 7l: catalog#42504606211, lot#64900906; persona revision str spln stem 16x135: catalog#42560113516, lot#64853724; persona stem extension straight splined uncemented 15 mm diameter +135 mm length: catalog#42560113515, lot#65042462; persona articular surface fixed bearing constrained posterior stabilized (cps) left 16 mm height: catalog#42512600516, lot#64099963; unknown stryker cement: catalog#ni, lot#ni.The product will not be returned to zimmer biomet for investigation, as the product currently remains implanted.The investigation is in process.Upon receipt of additional information or completion of the investigation, a follow-up mdr be submitted.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d10; g3; h2; h3; h4; h6; h10; h11.D10: medical product: stem extension straight splined uncemented 15 mm diameter +135 mm length: catalog#42560113515, lot#65042452.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Regarding the reported pain, root cause was unable to be determined.Regarding the reported tendonitis and bursitis, the root cause of the reported event was determined to be not related to the device.Bursitis is the inflammation or irritation of the bursae (the fluid filled sac that cushions the joint) and is typically caused by repetitive motion, overuse, and pressure to the bursae.Bursitis is a very common condition that can impact any of the joints and can last for a short duration or years.Conservative treatment consists of over the counter (otc) medications pain relievers and anti-inflammatories, rest, ice, elevation, and applying pressure wraps.If conservative treatments fail, physical therapy, aspiration, arthroscopy, or steroid injections may be necessary.The complaint indicates that postop bursitis developed and required medical intervention for treatment.Tendonitis is the inflammation or irritation of the tendons and is typically caused by repetitive motion, overuse, and pressure to the bursae.Patients can experience a decrease in overall adls, rom of the joint, decrease in quality of life, as well as an increased need for possible over-the-counter (otc) medications for swelling and pain control.This condition can impact the patient's ability to use the joint fully, decrease overall activities of daily living (adls), range of motion (rom) of the joint, and quality of life, while increasing the need for possible over-the-counter (otc) medications for swelling and pain control.Treatment options include otc medications, prescribed pain medications, physical therapy, rest, ice, elevation of the affected joint, and steroid injections.Reported event was confirmed by review of provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIA FIXED CEMENTED LEFT SIZE D STEM EXTENSION USE REQUIRED
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22472140
Report Number0001822565-2025-02455
Device Sequence Number18106201
Product Code JWH
UDI-Device Identifier00889024562783
UDI-Public(01)00889024562783(17)3107020(10)65069838
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K191625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number42542006701
Device Lot Number65069838
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2025
Supplement Date Received by Manufacturer12/15/2025
Initial Report FDA Received Date07/10/2025
Supplement Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.; SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight91 KG
Patient RaceWhite
-
-