The alleged event is a risk associated with breast surgery and there is no evidence to suggest a link between this particular implant and/or its manufacturing process and a higher risk of occurrence.The cause of malposition is multifactorial, and we cannot conclude that the reported event was caused by the manufacturing process and/or the product itself.A complete review of the dhr for lot involved was carried out, no deviation was found in the manufacturing process.Additionally, there were no indications of abnormalities in raw materials or manufacturing processes which may have affected this particular batch.The instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: malposition: malposition of a breast implant is defined as an incorrect placement during surgery or its shifting from its original position.It is also called displacement/lateralization.Malposition has been a frequent event reported due to its multifactorial causes and it can be expected during the lifetime of the device.The implant¿s shifting can be produced by trauma, capsular contracture, gravity, or initially wrong placement.The surgeon must plan the operation carefully and conduct the surgery with a technique that can minimize but not completely evade the risk of malposition.The risk associated with this event is dissatisfaction with aesthetic outcomes.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health establishment labs will continue to monitor for similar complaints/trends.As part of the post market surveillance process, monitoring of the primary endpoints reported is performed to determine unfavorable trends.Per our current complaint data report, no unfavorable trends were detected on this product.No further actions are required.
|
|
General conclusion device involvement: a causal relationship between the reported event and the device has been established.Event characterization: the event was classified as malposition,seroma,pain.Clinical confirmation upon thorough evaluation of the submitted clinical documentation and supporting evidence, the reported case was confirmed.Root cause analysis this event is a well-documented complication inherent to breast implant surgery.Based on the analysis of the available data, including clinical findings and product traceability records, there is no evidence indicating a causal link between this specific case and any product-related factors, including raw materials or manufacturing processes.The etiology of the alleged event is recognized as multifactorial, given the absence of manufacturing deviations or anomalies, and in line with current clinical literature, the event is considered not attributable to the manufacturing process and/or the product itself.Dfu and labeling evaluation the event is a known, well-documented complication associated with silicone breast implants.It is clearly addressed in the product¿s directions for use (dfu), which states: seroma ¿ seroma is an accumulation of fluid that results from tissue inflammation7.The etiology of seroma is known in breast surgery and is connected to a hypovascular milieu or trauma following the surgery.Often, seromas are reabsorbed by the body over several weeks, but needle drainage is sometimes needed to remove the fluid8.While seromas do not increase breast cancer risk, they sometimes heal with scar tissue or calcifications that can raise a concern about mammograms in the future.Seroma symptoms most often appear a week to 10 days after surgery; the area may feel tender and swollen, with a discrete lump and redness arising within a day or two.Early seroma formation is defined as periprosthetic fluid accumulation within the first postoperative year, whereas the late form is any moment beyond that time9.In addition to causing pain, a seroma increases the risk of developing an infection in the breast.Depending on the location, it may also increase pressure over the surgical site and sometimes cause wound dehiscence.Pain ¿ most women undergoing augmentation or reconstruction with a mammary implant will experience post-operatory breast and/or chest pain.While this pain usually recedes in most women as they heal after surgery, it can become a chronic problem in other women.Hematoma, migration, infection, implants that are too large or capsular contracture can cause chronic pain.Sudden, severe pain may be associated with implant rupture.The surgeon should instruct the patient to immediately report if there is significant pain or if pain persists.Malposition ¿ malposition of a breast implant is defined as an incorrect placement during surgery or its shifting from its original position.It is also called displacement/lateralization.Malposition has been a frequent event reported due to its multifactorial causes and it can be expected during the lifetime of the device.The implant¿s shifting can be produced by trauma, capsular contracture, gravity, or initially wrong placement16.The surgeon must plan the operation carefully and conduct the surgery with a technique that can minimize but not completely evade the risk of malposition.The risk associated with this event is dissatisfaction with aesthetic outcomes.The dfu also emphasizes the responsibility of the surgeon to fully inform the patient of potential risks and complications during the pre-operative consultation.Patients are provided with the document ¿motiva implants®: information for the patient,¿ accessible at ifu.Motiva.Health.Literature reference: seroma ¿late seroma is defined as a periprosthetic fluid collection occurring more than 1 year following breast augmentation¿ (nava, et al.2017).¿early seroma formation is defined as periprosthetic fluid accumulation within the first postoperative year, whereas the late form is any moment beyond that time.¿ (sforza, et al.2017) ¿seroma can present both early and late, with the latter being recently linked with malignant growths and prolonged onset.¿ (sforza, et al.2017) some analysis have revealed that the pocket plays a significant role in developing seroma, patients with submammary pocket developed seroma in higher rates when compared with patients with submuscular pocket.Submammary pocket increases the odds of developing seroma by 7.5 times.(sforza, et al.2017) some factors are significantly associated with seroma development: a high bmi, large implant size, submammary pocket, and smoking which significantly amplified the effects of other variables.(m.Sforza, et al., 2017 unraveling factors influencing early seroma formation in breast augmentation surgery.Aesthetic surgery journal 2017, vol 37(3) 301¿307 © 2016 the american society for aesthetic plastic surgery, inc.Doi: 10.1093/asj/sjw196.Larger implant size are also known to be related to higher incidences of other complications such as capsular contracture and hematoma.(collins jb, verheyden cn.Incidence of breast hematoma after placement of breast prostheses.Plast reconstr surg.2012;129(3):413e-420e) malposition karan chopra, md, arvind u.Gowda, md, edwin kwon, md, michelle eagan, md, w.Grant stevens, md, techniques to repair implant malposition after breast augmentation: a review, aesthetic surgery journal, volume 36, issue 6, june 2016, pages 660¿671, https://doi.Org/10.1093/asj/sjv261.Device history record (dhr) review a comprehensive review of the device history record (dhr) for the affected lot was completed.No deviations, non-conformances, or anomalies were identified during manufacturing.All raw materials and process parameters conformed to specified quality requirements.Trend and signal monitoring the event has been evaluated as part of ongoing trend analysis activities within the pms program.Per the current complaint data report: no adverse or unexpected trends related have been identified.No further corrective or preventive actions are required at this time.Establishment labs will continue to monitor for similar events as part of routine pms surveillance.
|