• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPACELABS HEALTHCARE XHIBIT TELEMETRY RECEIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE XHIBIT TELEMETRY RECEIVER Back to Search Results
Model Number 96280
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported a few telemetry boxes would drop off the xhibit central stations (xcs) intermittently.The telemetry patients would return on their own, however customer settings and notes were set to default.A field service engineer (fse) went on site and confirmed the system was operational.An error message was seen in the xcs message bar indicates the xhibit telemetry receiver (xtr) had disconnected and reconnected.As a precaution the fse gathered device logs and forwarded them to the product support specialist (pss) for additional investigation.Apss reviewed the xcs logs and confirmed that a xtr had disconnected on 9/jun/2025 for roughly 30 seconds.Xtr logs were unavailable for review.A pss made several attempts to reach out to the customer to gather additional data regarding the loss of telemetry monitoring and access to the xtr logs, however, the customer has not responded.Should additional information be made available, a supplemental report will be submitted in accordance with 21 cfr 803.56.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XHIBIT TELEMETRY RECEIVER
Common Device Name
TELEMETRY RECEIVER
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key22476005
Report Number3010157426-2025-00115
Device Sequence Number8668850
Product Code MHX
UDI-Device Identifier10841522100246
UDI-Public(01)10841522100246
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K141156
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2025
Initial Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-