C.R. BARD, INC. (BASD) -3006260740 POWERPORT MRI 8GS, INT. W SP, ATT S/L; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
1878000 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Migration (4003)
|
| Health Effect - Clinical Codes |
Nausea (1970); Pain (1994); Tachycardia (2095); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient underwent for port placement procedure using powerport mri 8gs, int.W sp, att s/l.It was reported that during a port placement procedure, the port was allegedly found fractured.It was further reported that the patient had a cardiac cath to remove the fractured piece from the atrium and ventricle of the heart.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) are adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos records were provided for review.Therefore, the investigation is inconclusive for the reported catheter fracture, separation and migration issues as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, d4 (unique identifier (udi) #), g3, h6 (patient, component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
A patient underwent for port placement procedure using powerport mri 8gs, int.W sp, att s/l.It was reported that during a port placement procedure, the port was allegedly found fractured.It was further reported that the patient had a cardiac cath to remove the fractured piece from the atrium and ventricle of the heart.Reportedly, patient allegedly experienced tachycardia, pain, nausea.The current status of the patient was unknown.
|
| |
|