• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORBUSNEICH MEDICAL (SHENZHEN) CO., LTD JADE 18 OTW PTA BALLOON DILATATION CATHETER; Catheter, angioplasty, peripheral, transluminal

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORBUSNEICH MEDICAL (SHENZHEN) CO., LTD JADE 18 OTW PTA BALLOON DILATATION CATHETER; Catheter, angioplasty, peripheral, transluminal Back to Search Results
Catalog Number 597010022
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that non-abbott laser was used for atherectomy to treat a diseased superficial femoral artery (sfa).After atherectomy, a stent was placed, post stent dilation was performed with a jade.018 otw 7.00mmx100mm 150 cm balloon.The balloon was dilated up to burst pressure of 16 atmospheres (atm), and the balloon burst.The physician went double negative on the balloon, pulled the balloon back to take out of body.When balloon was removed from the body, it was evident on the catheter that only a piece of the balloon was still attached.After fluoroscopy, remanent of the balloon appeared still in the patient.The physician attempted to retrieve the balloon within another catheter and was unsuccessful.The patient left for surgery.A cutdown was performed to remove the balloon fragments.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JADE 18 OTW PTA BALLOON DILATATION CATHETER
Common Device Name
Catheter, angioplasty, peripheral, transluminal
Manufacturer (Section D)
ORBUSNEICH MEDICAL (SHENZHEN) CO., LTD
no.1st jinkui road
futian free trade zone
shenzhen, guangdong 51803 8
CH  518038
MDR Report Key22477545
Report Number3014148220-2025-00243
Device Sequence Number8040379
Product Code LIT
UDI-Device Identifier06934955904028
UDI-Public(01)06934955904028(11)231010(17)251010(10)5948352309
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number597010022
Device Lot Number5948352309
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/20/2025
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/11/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-