| Model Number |
SY60WF |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-health care professional reported that following an intraocular lens (iol) implant procedure, at one week post operative visit, physician noticed haptic was no longer attached to the iol.The iol was explanted and exchanged with an unspecified iol in the secondary implant procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been provided in h.3.And h.11.The product was not returned.A photo was provided of a monitor screen.The lens was shown inside the eye, anterior surface up.One haptic was broken in the distal area.The broken haptic was positioned onto the anterior optic surface.A metal instrument was shown being inserted a the optic edge.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Associated products were not provided.It is unknown if qualified products were used.Based on our observation of the attached photo, the haptic is broken.Information was provided that the broken haptic was noticed one week post-op.The account indicated the product was not available to evaluate.It is difficult to make a final determination without evaluation of the physical sample.The root cause cannot be determined based on available information.Associated products were not provided.It is unknown if qualified products were used.The ifu instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|