BAUSCH + LOMB BAUSCH + LOMB INJECTOR SYSTEM CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number BLIS-X1
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code Zonular Dehiscence (2698)
Date of Event 05/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As per the reporter, the device is not available for analysis.As a lot number for the device was not provided, the associated device history record was not reviewed.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the information provided, a root cause of this event could not be determined.
 
Event or Problem Description
It was reported that during implantation of an intraocular lens (iol) into the right eye, the lens unfolded too quickly causing zonular damage and iol decentration.The incision was enlarged to allow for removal of the lens, a different model iol was implanted into the sulcus, and the incision was closed with sutures.In the surgeon¿s opinion, the most likely cause of the event was quick unfold of lens after implant.Patient outcome is good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAUSCH + LOMB INJECTOR SYSTEM CARTRIDGE
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
medicel ag
dornierstrasse 11
altenrhein CH ¿ 9423
SZ   CH ¿ 9423
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key22478155
Report Number0001313525-2025-00146
Device Sequence Number8554778
Product Code MSS
UDI-Device Identifier00757770520183
UDI-Public(01)00757770520183
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131958
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBLIS-X1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2025
Initial Report FDA Received Date07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVISTA TORIC INTRAOCULAR LENS, DUOVISC.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-