• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number EE2110X15B
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a broken fan for a trilogy evo device.There was no harm or injury reported.The device was not in patient use.During the evaluation of the device at the third-party service center, a ventilator inoperative error was discovered in the device's event log.The technician recommended replacing the device's circuit board to address the issue.No further investigation is required at this time.Additionally, the device's in-line filter is contaminated, air inlet foam filter replace due to preventive maintenance and were replaced.The device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22478704
Report Number2518422-2025-107269
Device Sequence Number15463937
Product Code CBK
UDI-Device Identifier00606959055544
UDI-Public(01)00606959055544
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEE2110X15B
Device Catalogue NumberEE2110X15B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2025
Initial Report FDA Received Date07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-