This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, h10, and h11.Review of the most recent repair record determined the overheating issue was confirmed as the unit was smoking.Display would not power on, and the display cable had burn marks on.The cart board, display cable and the power supply were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|