| Model Number |
G6626525 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal therapy.It was reported that the cage was broken when trying to insert screws.There were no broken fragments left inside the patient and no further complications or symptoms reported regarding the event.Additional information was received via manufacturer representative that there is no allegation or malfunction associated with screw.The procedure involved in the event is anterior cervical discectomy and fusion (acdf) and pre-op diagnosis is cervical pain.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis of part# g6626525, lot# 73rv visual and optical inspection revealed the implant was damaged.The implant has cracked next to where the release/lock mechanism is.The implant is fragile and can be overloaded.It appears the implants structural integrity was overloaded during the implantation procedure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|