Model Number 0286
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
This implantable cardioverter defibrillator (icd) system exhibited elevated and fluctuating right ventricular (rv) lead impedance following a generator replacement.Historical data showed a gradual rise in impedance prior to replacement, followed by an expected drop with the new device, then minor fluctuations.No pacing inhibition or rv lead noise was observed.Due to the variability, technical services recommended further in-office evaluation to rule out under-insertion or intermittent lead fracture.Slightly elevated power consumption was attributed to persistent at/af and high rv sensing rate.The device is depleting normally, no malfunction was identified, and no adverse patient effects were reported.The system remains in service.
 
Event or Problem Description
This implantable cardioverter defibrillator (icd) system exhibited elevated and fluctuating right ventricular (rv) lead impedance following a generator replacement.Historical data showed a gradual rise in impedance prior to replacement, followed by an expected drop with the new device, then minor fluctuations.No pacing inhibition or rv lead noise was observed.Due to the variability, technical services recommended further in office evaluation to rule out under insertion or intermittent lead fracture.Slightly elevated power consumption was attributed to persistent at/af and high rv sensing rate.The device is depleting normally, no malfunction was identified, and no adverse patient effects were reported.The system remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22480264
Report Number2124215-2025-46465
Device Sequence Number20957014
Product Code NVY
UDI-Device Identifier00802526431418
UDI-Public(01)00802526431418(17)151204
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/04/2015
Device Model Number0286
Device Catalogue Number0286
Device Lot Number105316
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer07/17/2025
Initial Report FDA Received Date07/11/2025
Supplement Report FDA Received Date07/18/2025
Date Device Manufactured12/04/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-