• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING (CR) NARROW POROUS; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING (CR) NARROW POROUS; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42502206602
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 02/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: unknown persona articular surface.Unknown persona tibial component.G2: literature citation: c.Nakasone, i.Weber, c.Israelite et al., (2025).Early radiographic evaluation of an anatomic porous tantalum tibia: a prospective, multi-center, non-randomized clinical study.The knee (53), 264-272.Https://doi.Org/10.1016/j.Knee.2025.01.010.H6: proposed component code: mechanical (g04) - femur.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported through a journal article and study that one patient underwent manipulation under anesthesia due to limited range of motion approximately 6 weeks post implantation.Attempts for additional information are in progress and none has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b3, b4, b5, d1, d2b, d4, d6a, d10, e1, g3, h2, h6.D10: 42530007102, tibia trabecular metal two-peg porous fixed bearing right size e, 62810182.42522200512, articular surface fixed bearing ultracongruent (uc) right 12 mm height, lot 64016050.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was further reported that approximately 2 months post implantation of a knee arthroplasty, the patient underwent a manipulation under anesthesia due to limited range of motion.Attempts for additional information have been made and none is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, h2, h6.The root cause of the reported event is not device related.Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known postoperative procedure related complication that can occur from surgical implantation of new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA FEMUR TRABECULAR METAL CRUCIATE RETAINING (CR) NARROW POROUS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22480652
Report Number0001822565-2025-02475
Device Sequence Number7942602
Product Code MBH
UDI-Device Identifier00889024470125
UDI-Public(01)00889024470125(17)271231(10)63807056
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
NI
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number42502206602
Device Lot Number63807056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Supplement Date Received by Manufacturer07/11/2025
08/14/2025
Initial Report FDA Received Date07/11/2025
Supplement Report FDA Received Date08/05/2025
08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE; SEE H11 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-