• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121732048
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Ambulation or Postural Difficulties (2544)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: a search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The product was not returned to j&j medtech orthopaedics; however, photos were received for review.The x-ray investigation revealed that a disassociation occurred between the acetabular liner and the cup.The femoral head appears to be not centrally loaded since it can be observed that it is having a direct contact with the cup.It is suspected an audible sound condition can happen as a consequence of the disassociation.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the cup is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.A search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The overall complaint was confirmed as the observed condition of the pinn sector w/gription 48mm would have contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported that patient underwent revision surgery due to worn implants.There was no surgical delay.On (b)(6) 2024, the patient underwent a successful left total hip arthroplasty using depuy synthes implants.On (b)(6) 2025, the patient has a follow-up appointment that states a few days prior, the patient noticed a squeaking sound following bending and increased flexion of the left hip.The patient also has pain when loading the joint.The patient limps and has an antalgic gait pattern.X-rays show the insert has wear on the left side causing superior migration of the head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR W/GRIPTION 48MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22480793
Report Number1818910-2025-11895
Device Sequence Number18128322
Product Code KWA
UDI-Device Identifier10603295010289
UDI-Public(01)10603295010289(17)20331231(10)4355374
Combination Product (Y/N)N
Initial Reporter CountryHU
PMA/510(K) Number
K093646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121732048
Device Lot Number4355374
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/09/2025
Initial Report FDA Received Date07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORAIL AMT COLLAR SIZE 11.; DELTA CER HEAD 12/14 32MM +1.