• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number TLEAD1058-70B
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient experienced a trial lead breakage in half.An hcp assessment regarding the cause of the reported event was not available.The patient had the trial device explanted.There have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
H1.Only one option can be selected.However, this case is both a malfunction and serious injury.A review of the device's manufacturing records was completed, including batch history and device history; there were no deviations or nonconformities associated with the reported event.Nevro is awaiting the return of the device.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Additional Manufacturer Narrative
The lead was returned and analyzed.Visual inspection of the returned lead confirmed the reported complaint about the lead being fractured.Optical microscopy showed the lead fracture likely occurred from being cut and not from stress, strain, or plastic deformation.However, the exact root cause of the fracture could not be determined as the damage could have been procedure, hcp, or patient related.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
john paxton
1800 bridge parkway
redwood city, CA 94065
6504332730
MDR Report Key22482118
Report Number3008514029-2025-00543
Device Sequence Number8668654
Product Code LGW
UDI-Device Identifier00813426021104
UDI-Public00813426021104
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTLEAD1058-70B
Device Lot Number94585818
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Supplement Date Received by Manufacturer09/02/2025
Initial Report FDA Received Date07/11/2025
Supplement Report FDA Received Date09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TLEAD1058-50B LN 94586072.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-