| Catalog Number |
EN2918 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, during implantation there was an aneurysm on both narrow cylinders on test inflation.It started after mild modeling.A new device was used to successfully complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump and cylinders 1 and 2 were received for evaluation.No functional abnormalities were noted with the pump.An aneurysm was noted on the bladder of cylinder 1.This was not a site of leakage.No functional abnormalities were noted with cylinder 2.Based on the procedures by which this device is tested prior to release, specific to revealing an aneurysm, it was concluded this most likely occurred subsequent to the opening of the device packaging.In addition this most likely occurred as the result of overinflation and/or excessive manipulation during the attempt to correct the observed penile curvature.This device is capable of generating pressure sufficient to aneuryze an unrestricted bladder.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, during implantation there was an aneurysm on both narrow cylinders on test inflation.It started after mild modeling.A new device was used to successfully complete the procedure.
|
| |
|
Search Alerts/Recalls
|