| Model Number |
AIQCA |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned as it was discarded by the hospital.However, a supplemental report will be forthcoming once the engineering evaluation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|
Event or Problem Description
|
|
It was reported that after use of the acumen iq finger cuff, positioned adjacent to a finger with a pulse oximetry sensor, the patient developed a blister on their finger tip.The device is not available for evaluation as it was disposed.There was no other patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Since the complaint affected unit was not returned for evaluation, a product non-conformance or device failure associated to manufacturing or design could not be confirmed.The lot number was not provided thus a device history record was not reviewed.The instructions for use contain the following warnings: do not apply the finger cuff on injured skin as this may cause further injury.To reduce the risk of skin irritation and tissue damage, do not monitor longer than 8 hours continuously on a single finger.To continue to monitor beyond 8 hours, use an additional finger cuff on another finger or move the cuff in use to another finger.
|
| |
|
Search Alerts/Recalls
|