| Model Number |
V+C+S+N+PRO, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to ventec that the device needed service.During the device's evaluation, ventec observed that it was providing low fio2 (fraction of inspired oxygen) readings.There were no reports of patient use associated with the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of it providing low fio2 (fraction of inspired oxygen) readings was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ift.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information for supplemental medwatch report 001: section d9, date returned to mfg, of the initial medwatch report stated: blank.Section d9, date returned to mfg, of the initial medwatch report should have stated: 03/04/2025.
|
| |
|