The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the cmos module with drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the control body corroded, the bending rubber leak, the bending rubber cut, the led cover glass signs of moisture or fluid under and the cover glass; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586 (image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|