ARTHROCARE CORPORATION QFIX 1.8 MINI SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72290123 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, after the insertion site was cleared of all debris and cycled multiple times with a drill, the q-fix sutures snapped inside the inserter when the surgeon attempted to pull it out.The anchor remained inside the patient and was not removed, and the sutures were cut.Surgery was completed using a smith and nephew backup device instead, used in the originally drilled bone hole.There was no surgical delay, and no further complications were reported.The patient is recovering post-surgery.
|
| |
|
Search Alerts/Recalls
|
|
|