C.R. BARD, INC. (BASD) -3006260740 PowerPort ClearVUE Slim Implantable Port, ChronoFlex Single-Lumen, Kit, 8F; Port & Catheter, Implanted, subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
5616000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Purulent Discharge (1812); Erythema (1840); Unspecified Infection (1930)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.Therefore, the investigation is inconclusive for the reported infection as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that sometime later port placement procedure, the patient developed with unspecified infection.The current status of the patient was unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through the litigation process that, on (b)(6)2022, a patient underwent port placement via the left subclavian vein for chemotherapy related to colon cancer.Approximately one month and twenty nine days later, on (b)(6) 2023, the patient presented with erythema and drainage over the left chest at the port site.Approximately one day later, on (b)(6) 2023, the patient was diagnosed with a bacterial infection, which was confirmed by port culture.Subsequently, the port was removed on the same day.Additionally, approximately three days later, on (b)(6)2023, the patient was diagnosed with a bacterial infection due to micrococcus and gram negative rods.It was further reported that the patient eventually expired.However, the cause of death was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information was identified in medical records.Based on the review of the medical record, and the file was reassessed for reportability and determined to be reportable as death.The date of death for this patient was not provided; therefore, the date of death has been updated as 01-jan-1900 to meet the mdr submission requirement.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.B2, b3, b5, g3, h1, h6 (patient, result, conclusion)
section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|