|
Updated fields: b4, b5, b6, b7, d8, d9, d10, g1 (contact person mfg site), g3, g6, h2, h3, h6 (type of investigation, investigation findings, health effect impact codes and investigation conclusions), h11.Corrected field: h6 (medical device problem code).A getinge field service engineer (fse) was dispatched to evaluate the unit.Fse states, the equipment was checked, and it was displaying the following message "bbia may need maintenance." the error logs were checked for details on the message in question.The equipment displayed error #37 fill fail - shuttle purge timeout in the logs, so several verification tests were performed.During the functionality tests, it did not displayed any error, but for safety, atmospheric calibration was performed again.The equipment was under testing for approximately 2 hours, while during this period it did not display the error message again and there were no failures.The equipment is compliant and released for use.
|