| Catalog Number |
8065751510 |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
06/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during cataract surgery, a burr was observed on the ophthalmic irrigation/aspiration tip during posterior chamber polishing, and the patient experienced a posterior capsule rupture.The surgery was completed on the same day after the product was replaced.The current condition of patient is resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
One opened polymer irrigation and aspiration (i/a) tip in a wrench, in a blister was received for the reported event.The returned sample was visually inspected and found to be conforming.Although a mold mismatch was observed, there were no sharp edges.A dimensional inspection was performed and found to be conforming with the mold mismatch observed to be within specification limits.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is not determined as the returned sample was conforming for all visual and dimensional testing.No further investigative or manufacturing actions are warranted at this time due to that the returned sample met specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|