SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 52MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
| Catalog Number |
74120152 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); High metal ion levels (4847)
|
| Date of Event |
09/20/2006
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was implanted with a right bhr-tha on (b)(6) 2006.The patient later presented with elevated metal ion levels (cr 40.9 ¿g/l, co 69.2 ¿g/l) and osteolysis.Imaging showed a retroverted femoral anteversion angle of -22°, which may have contributed to implant wear.No revision or surgical intervention was reported for the hip.The patient impact was limited to elevated metal ions and osteolysis.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5 (occurrence date is unknown), h6 (type of investigation, investigation findings).Updated results of investigation: as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for cup.No similar complaints have been identified for the head.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.There is a likely route cause of the elevated metal ions (cr 40.9 and co 69.2) and osteolysis.The retroverted femoral anteversion of -22° cannot be ruled out as a contributing factor as mal positioned implants can lead to accelerated wear of the implants, elevated metal ions and osteolysis.The patient impact is determined to be the osteolysis and elevated metal ions.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|
| |
|