STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SILVERGLIDE BIPOLAR STANDARD FORCEPS CUSHING, 19CM / 1.2MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 6710190012
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer service technician reported the silver coating was flaked off both tips of the device, posing the risk of material being lost in a surgical site.There were no adverse consequences related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILVERGLIDE BIPOLAR STANDARD FORCEPS CUSHING, 19CM / 1.2MM TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key22493451
Report Number3015967359-2025-01255
Device Sequence Number4121257
Product Code GEI
UDI-Device Identifier04546540532084
UDI-Public(01)04546540532084(10)391262
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K992931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number6710190012
Device Lot Number391262
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2025
Initial Report FDA Received Date07/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-