MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK REVISION TOTAL KNEE SYSTEM; GMK-SPHERE FIXED TIBIAL INSERT SC SIZE 4/12MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK REVISION TOTAL KNEE SYSTEM; GMK-SPHERE FIXED TIBIAL INSERT SC SIZE 4/12MM Back to Search Results
Model Number 02.07.0412SCF
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 06/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
At about 1 month after the primary, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout and revised the liner.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 02 july 2025, lot 2400347: (b)(4) items manufactured and released on 15-march-2024.Expiration date: 2029-02-24.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported events during the period of review.Infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK REVISION TOTAL KNEE SYSTEM
Common Device Name
GMK-SPHERE FIXED TIBIAL INSERT SC SIZE 4/12MM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key22493703
Report Number3005180920-2025-00675
Device Sequence Number4234352
Product Code JWH
UDI-Device Identifier07630030817755
UDI-Public07630030817755
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K103170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.07.0412SCF
Device Lot Number2400347
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2025
Initial Report FDA Received Date07/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
Patient Weight115 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-