• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC UV ABSORBING IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number CC60WF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported with a description of intraocular lens haptic was torn.Additional information has been received and stated that the lens was removed during initial procedure and replaced with another lens.The surgery was completed on the same day.There was no patient harm.
 
Additional Manufacturer Narrative
Correction information was provided in d.10.Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned outside of the fully disassembled lens case, loosed inside a plastic bag.Solution was dried on the lens.One haptic was broken or torn in the gusset area (returned).The optic was cut in half, typical of insertion and removal.Product history records were reviewed and the documentation indicated the product met release criteria.Per the questionnaire the associated products were unknown.It is unknown if qualified products were used.The reported haptic damage was observed.All lenses are 100percentage inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.It is unknown if a qualified associated products were used.The ifu instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturers recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.No further information has been provided at this time.File will be reopened if new information is received.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22494422
Report Number1119421-2025-01925
Device Sequence Number4097060
Product Code HQL
UDI-Device Identifier00380652396868
UDI-Public(01)00380652396868(17)290205(11)250207(10)15986746
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCC60WF
Device Lot Number15986746
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/19/2025
Supplement Date Received by Manufacturer07/15/2025
Initial Report FDA Received Date07/14/2025
Supplement Report FDA Received Date08/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN CARTRIDGE; UNKNOWN HANDPIECE; UNKNOWN VISCOELASTIC
Patient Age66 YR
Patient SexMale
-
-