|
|
| Model Number |
CC60WF |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned positioned correctly in the lens case.Solution was dried on the lens.The optic was torn, split, and cracked almost into two pieces, similar in appearance to damage from insertion and removal.The lens was cleaned with klrp for further evaluation.The posterior of the optic had a line of scrape marks.Product history records were reviewed, and documentation indicated the product met release criteria.Qualified associated products were indicated.The posterior optic had scrape marks in a line, typical in appearance to damage from an instrument used to insert the lens into a cartridge and similar to damage from a handpiece plunger during a plunger underride of the lens.This optic damage was most likely interpreted as the reported complaint.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The instructions for use (ifu) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs to completely fill the cartridge with ophthalmic viscosurgical devices (ovd) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company intraocular lens (iol) to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The company 29.0 diopter lens was removed and replaced with another company 29.0 diopter lens.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, after insertion a defect was noted in the center of the lens, so doctor opted to remove and replace.The procedure was completed on the same day with company same lens model and diopter.There was no patient harm.In the surgeon¿s opinion defect in the cartridge contributed to event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The lens was returned positioned correctly in the lens case.Solution was dried on the lens.The optic was torn, split, and cracked almost into two pieces, similar in appearance to damage from insertion and removal.The lens was cleaned with klrp for further evaluation.The posterior of the optic had a line of scrape marks.Product history records were reviewed, and documentation indicated the product met release criteria.Qualified associated products were indicated.The posterior optic had scrape marks in a line, typical in appearance to damage from an instrument used to insert the lens into a cartridge and similar to damage from a handpiece plunger during a plunger underride of the lens.This optic damage was most likely interpreted as the reported complaint.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The instructions for use (ifu) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ophthalmic viscosurgical devices (ovd)-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The company 29.0 diopter lens was removed and replaced with company another 29.0 diopter lens.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|
|
|