| Catalog Number |
386801 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd cathena 18gx1.25in straight bc had foreign matter on assessment, the cannulas appear to have a clear residue or flake on the cutting edge (anterior bevel).This residue raises concerns about sterility and possibility to affect the effectiveness of the cannulas during use.Action: actions taken include conducting a thorough inspection of the affected cannulas to confirm the presence of residue on the anterior bevel.The findings were recorded, including product details, batch numbers, and photographs of the affected cannulas.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Our quality engineer inspected the 2 samples submitted for evaluation.The reported issue of foreign matter was confirmed upon inspection of the samples.Analysis of the samples showed foreign matter on the cannula tip and cover.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records have been reviewed, and this batch meets our manufacturing specification requirements.The assembly process of the cannula lubrication process was reviewed.The cannula is lubricated by dipping it into a silicone lube tank.During dipping, low air flow is blown through the cannula to prevent silicone flowing into the cannula.After dipping, the part is raised from the silicone lube tank followed by high pressure air blown through the cannula to remove the excess silicone.The catheter subassemblies are then set together with the needle hub subassembly.The assembled part then goes through a series of processes and loaded with the needle cover.There could have been silicone accumulated at the surface of the lube tank which might have transferred to the cannula surface during air blowing.After the catheter subassemblies are set together with the needle hub subassembly, given the silicone-like nature of the lube it is possible for the excess silicone to migrate to the cannula/catheter tip area during transportation or storage.To prevent this defect, revision of the procedure will be completed to include cleaning of the surface of the lube tank and surrounding areas every shift.
|
| |
|
Search Alerts/Recalls
|