MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, bent haptic.There was no patient contact.The procedure was completed on the same day with another lens.Additional information was requested.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.The lens was returned positioned incorrectly in a lens case.Solution was dried on the lens.Both haptics were bent in the gusset and distal area.The optic was torn against and post and pressed down into the well area of the lens case.Product history records were reviewed, and the documentation indicated the product met release criteria.Qualified associated products were indicated.Upon return, the lens was observed to be placed into the lens case incorrectly resulting in damage.The reported bent haptic was observed.Other lens damage was observed that may have occurred due to how the lens was positioned in the lens case for return.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company's release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The ifu (instructions for use) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd (ophthalmic viscosurgical device)-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol (intraocular lens) to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22495706
Report Number1119421-2025-01929
Device Sequence Number8477637
Product Code HQL
UDI-Device Identifier00380652251433
UDI-Public(01)00380652251433(17)280909(11)240911(10)15915631
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.175
Device Lot Number15915631
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2025
Supplement Date Received by Manufacturer07/25/2025
Initial Report FDA Received Date07/14/2025
Supplement Report FDA Received Date08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II LOADING FORCEPS.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE.
Patient SexUnknown
-
-