Catalog Number 136521000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 06/26/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that surgeon had a patient presented with a disassociated pinnacle altrx liner requiring revision.There was loosening of 48x32 n altrx liner at non cemented interface.Cement manufacture was unknown.The femoral head was displaced superior and rubbing the superior rim of the cup.After explanting the femoral head and remaining liner, thorough washout/debridement, and cup/stem stability/fixation checked, a dm construct was determined to best correct possible causations.Hip was reduced with a 48x41 dm liner, 41x22.225 head liner, and 22.225 metal head.Stability was satisfactory.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown; therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICUL/EZE BALL 32 +1 GR
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds LS11 8DT
UK   LS11 8DT
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22495989
Report Number1818910-2025-12005
Device Sequence Number12915287
Product Code JDI
UDI-Device Identifier10603295033172
UDI-Public(01)10603295033172
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K883460
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number136521000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/26/2025
Supplement Date Received by Manufacturer07/18/2025
07/22/2025
Initial Report FDA Received Date07/14/2025
Supplement Report FDA Received Date07/21/2025
07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 32IDX48OD; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-