|
MENTOR TEXAS MENTOR SMOOTH ROUND HIGH PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3503290 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
06/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 290cc mentor smooth round high profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with a deflated right breast implant during an in-office exam.As a result, the patient underwent bilateral exchange with mentor saline implants on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On august 05, 2025, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On august 22, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted a visual inspection, leak test, microscopic examination, and shell thickness of the device.During visual analysis of the returned device, no apparent damage or visual anomalies were observed.Leak testing was performed, in accordance with mentor's procedures.Findings revealed a tear on the anterior view, near the valve system.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: according to the product insert data sheet, the valve system can be damaged by improper use of the fill-tube stylet.Care should be taken that the stylet enters the valve smoothly.Use the thumb and forefinger to stabilize/support the valve seat and gently push the stylet tip into the valve opening.Overstressing the valve material may result in punctures or tears and subsequent deflation may occur.Use only the fill tube stylet provided with this product.Take care not to puncture the diaphragm valve or the shell with the stylet tip.Care must also be taken when the fill tube stylet is removed to prevent damage to the valve assembly.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|