| Model Number |
14000 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the patient's blood glucose levels (bg) rose to over 250 mg/dl, while wearing the pod between 4 and 24 hours.The pod reportedly fell off the infusion site (leg), causing the cannula to dislodge.No treatment information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the complainant the device will not be returned for investigation.It was reported that the cannula had dislodged from the infusion site.This condition could interrupt insulin delivery and contribute to hyperglycemia.No lot release records were reviewed, as the product lot number was not provided.
|
| |
|
Search Alerts/Recalls
|