| Model Number |
NA-403D-2021 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the subject device had an incomplete retraction issue during a diagnostic procedure.The procedure was completed using a similar device.There were no reports of patient harm or injury.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation: it is likely the following led to the malfunction: the most probable cause of the complaint was traced to component failure, which is expected or random component failure without any design or manufacturing issue.Deformation/distortion problem, problems caused by changes in the shape or size of the device due to an applied force.This can be a result of tensile forces (pulling), compressive forces, shear, bending, knotting, or torsion.No further escalation is required.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|