• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 NEUT 28IDX50OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 NEUT 28IDX50OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Catalog Number 121828650
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the reduction of the hip, the surgeon broke one of the poly slits.As a result, the femoral head could no longer lock into the liner.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Product description: pinn lnr con +4 neut 28idx50od.Product code:121828650.Lot no: m65h28.Quantity of manufactured: (b)(4).Date of manufacturing: 2024-06-07.Expiry date: 2029-05-31.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description: pinn lnr con +4 neut 28idx50od.Product code:121828650.Lot no: m65h28.Quantity of manufactured: (b)(4).Date of manufacturing: 2024-06-07.Expiry date: 2029-05-31.Device history review: a manufacturing record evaluation was performed for the finished device [lot/serial/batch] number, and no non-conformances / manufacturing irregularities were identified.Corrected: h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN LNR CON +4 NEUT 28IDX50OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22501063
Report Number1818910-2025-12033
Device Sequence Number7000550
Product Code KWZ
UDI-Device Identifier10603295011514
UDI-Public(01)10603295011514(17)20290531(10)M65H28
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121828650
Device Lot NumberM65H28
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2025
Supplement Date Received by Manufacturer07/22/2025
Initial Report FDA Received Date07/14/2025
Supplement Report FDA Received Date07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BONE CEMENT
Patient Age77 YR
Patient SexFemale
-
-