MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3545557MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
06/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast surgery with implantation of a 555cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured right breast implant using magnetic resonance imaging.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.Follow-ups have been conducted, and no additional information has been received.If more information becomes available, mentor will submit a supplemental report accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
On september 25, 2025, mentor was notified that the patient underwent bilateral exchange with allergan (non-mentor) implants on (b)(6) 2025.Date of explantation: (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On october 29, 2025, the mentor analysis lab received the suspect medical device for evaluation.On november 11, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted visual inspection of the device.Visual analysis of the returned device revealed that the breast implant had an area of siltex cracking on the posterior view.In addition, a tear was noted on the posterior view within the siltex cracking measuring approximately 1.8 cm.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Manufacturer¿s reference number: (b)(4).
|
| |
|